A method for the determination of AFB1 content in peanut oil was developed by HPLC-ESI MS/MS, and the detection conditions were optimized. The results showed that the optimal detection conditions were obtained as follows: ion source temperature 110?℃, capillary voltage 3 500 V, electrospray ionization (ESI+), sheath gas pressure 270 kPa, auxiliary gas pressure 104 kPa, magnetic gas flow rate 0.1 mL/min, with methanol as extraction solvent. Under these conditions, AFB1 had good linear relationship in the range of 1.00-11.00 ng/mL, and the limit of detection, average recovery rate and RSD were 0.3 μg/kg, 99.7% and 0.27% respectively. The method had the items of simpleness, rapidness, sensitivity and accuracy, and the ion pair scanning provided by ESI-MS could reduce the matrix interference to a great extent. |